AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  213
 
 
 
Calibration: Principles 
 
 
Calibration control 
Calibration control 
Depending on the gas analyzer version and equipment, there are three 
methods for controlling calibration: 
•  Manual calibration 
•  Automatic calibration 
•  Externally controlled calibration 
All analyzer modules can be calibrated using any of the three methods. 
Calibration start 
•  Manual calibration is started manually via the gas analyzer’s display and 
control unit. 
•  Automatic calibration is started at time intervals determined by the 
internal clock or by an external control signal or manually via the gas an-
alyzer’s display and control unit. 
•  Externally controlled calibration is triggered by an external control signal. 
Wait until the warm-up phase has ended 
Calibration should only be started after the warm-up phase. 
Analyzer module  Duration of the warm-up phase 
Caldos25 1.5 hours 
Caldos27  30/60 minutes for class 1/2 measurement ranges
 1)
 
Fidas24  ≤ 2 hours 
Fidas24 NMHC  ≤ 2 hours 
Limas11 IR  Approx. 2.5 hours 
Limas21 UV  Approx. 2.5 hours 
Limas21 HW  Approx. 4 hours 
Magnos206  ≤ 1 hour 
Magnos28  ≤ 1 hour 
Magnos27  2 to 4 hours 
Uras26  Approx. 0.5/2 hours without/with thermostat 
ZO23  Approx. 15 minutes 
1) See "AO2000 Series" data sheet for class details. 
Plausibility check during the calibration 
If during calibration the gas analyzer finds implausible values (e.g. if the 
span and zero values are equal), calibration is stopped and an error message 
is generated. The values stored for the last calibration remain in effect. 
Status signal 
The "Maintenance mode" status signal is set during calibration.