AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  285
 
 
 
Fidas24: Checking the combustion gas feed path in the gas 
analyzer for seal integrity 
CAUTION 
The seal integrity test described in this section requires special training 
and under some circumstances involves working with the gas analyzer 
open and powered up. Therefore, it must only be performed by qualified 
and specially trained personnel.   
If these conditions are not provided or the prescribed materials are not 
available, a seal integrity test must be carried out by ABB Service.   
Regular seal integrity check of the combustion gas path in the gas analyzer 
The gas analyzer must be in operation (flame on). 
1  Inspection of combustion gas feed path with positive pressure (com-
bustion gas inlet to combustion gas nozzle):   
With a leak detector (measurement principle thermal conductivity) sniff 
all the connection points. 
2  Inspection of the combustion gas feed path with negative pressure (in 
the detector, after the combustion gas nozzle):   
Connect zero gas at the sample gas inlet.   
Envelope all the connection points one after the other with a small gas 
cloud containing hydrocarbons (e.g. with coolant or test gas containing 
hydrocarbons or an acetone-soaked rag).   
Observe the measured value display while doing so; if there is a positive 
change in the measured value, the connection concerned is leaking. 
If there is a leak, shut down the gas analyzer 
If a leak has been established in the combustion gas feed path inside the gas 
analyzer, the gas analyzer must be put out of operation; it may under no 
circumstances be put into operation again. The cause of the leak must be 
determined and remedied by ABB Service.