110  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Analog output module: Electrical connections 
 
 
Pin Signal 
1 AO1+ 
2 AO1– 
3 AO2+ 
4 AO2– 
5 AO3+ 
6 AO3– 
7 AO4+ 
8 AO4– 
Analog outputs AO1 to AO4 
0/4 to 20 mA (pre-set ex works to 4 to 20 mA), common negative pole, elec-
trically isolated to ground, can be connected to ground as required, in this 
regard, max. gain compared to local protective ground potential 50 V, work-
ing resistance max. 750 Ω. Resolution 16 bit. The output signal may not be 
less than 0 mA. 
Design 
4-pole or 8-pole plug-in terminal strip. Please refer to the information about 
the requisite material (see page 116)! 
Terminal layout 
An analog output is allocated in the sequence of the sample components for 
each sample component. The sequence of the sample components is docu-
mented in the analyzer data sheet (see page 75) and on the identification 
plate (see page 74).