AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  291
 
 
 
Limas11 IR, Limas21 UV, Limas21 HW, Uras26: Measurement of 
calibration cells 
Definition 
The measurement of a calibration cell in the Limas11 IR, Limas21 UV,   
Limas21 HW and Uras26 analyzer modules means: 
Determining what calibration cell "deflection" is equivalent to the calibration 
reading with test gas. The "deflection" is stored as the "set point" of the 
calibration cell. 
When should calibration cells be measured? 
We recommend measuring the calibration cells once a year. 
We recommend measuring the calibration cells 
•  after end-point calibration of a sample component with test gas or 
•  after any change in measurement range limits or 
•  after a relinearization (see page 292). 
Before measuring the calibration cells 
Prior to measuring the calibration cells, the zero and end points of the ap-
plicable sample components must be calibrated with test gases. 
Test gas 
During calibration cell measurement the zero gas supply must be turned on. 
Menu path 
MENU → Maintenance/Test → Analyzer spec. adjustm. → 
Measure cal.cell