AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  267
 
 
 
Inspection and maintenance 
CAUTION 
The tasks described in this chapter require special training and under 
some circumstances involve working with the gas analyzer open and 
powered up. Therefore, they should only be performed by qualified and 
specially trained personnel. 
 
 
Inspection 
Normal operation of the gas analyzer 
In normal operation the measured values from the installed analyzer mod-
ules are shown on the display screen and the green "Power" LED is lit. 
Periodic checks 
 
Module, assembly  Function      Interval 
 
Sample gas   
flowmeter 
Caldos25  10 to 90 l/h,   
max. 200 l/h 
Regularly 
 
Caldos27  10 to 90 l/h         
Fidas24  80 to 100 l/h         
Fidas24 NMHC  80 to 100 l/h         
Limas11 IR  20 to 100 l/h         
Limas21 UV  20 to 100 l/h         
Limas21 HW  20 to 90 l/h         
Magnos206  30 to 90 l/h         
Magnos28  30 to 90 l/h         
Magnos27  20 to 90 l/h         
Uras26  20 to 100 l/h         
ZO23  5 to 10 l/h ± 0.2 l/h         
Pneumatic module 
disposable filter 
Staining Regularly 
 
Gas analyzer gas 
lines 
Seal integrity  Regularly 
 
Seals between doors 
and housing 
Contamination, foreign material  Before each 
closing