AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  235
 
 
 
Caldos27: Single-point calibration with standard gas 
Single-point calibration with standard gas 
For measuring ranges ≥ class 1 in the Caldos27 analyzer module, a sin-
gle-point calibration can be carried out with standard gas (standard gas 
calibration). This standard gas calibration is carried out exclusively as span 
calibration and causes an amplification correction. Safety-related measure-
ments are excluded from this technique. Depending on the measurement 
task involved, the zero and end points should be verified periodically (rec-
ommendation: once a year). 
Note: Classification of measuring ranges into class 1 and class 2 is specified in 
"Advance Optima AO2000 Series Continuous gas analyzers" data sheet. 
Measurement range for standard gas 
When the Caldos27 analyzer module is ordered with standard gas calibra-
tion, the factory-set measurement range for the standard gas is 
0 to 60,000 rTC (rTC = relative thermal conductivity). During basic calibra-
tion in the factory one standard gas is calibrated in this measurement range. 
Due to sensor tolerances the values of other standard gases can differ by up 
to 5 % from this scaling. 
Set points for standard gases 
 
N
2
 10,000 rTC 
Air 10,070 rTC 
Ar 7,200 rTC 
CO
2
 7,500 rTC 
CH
4
 14,000 rTC 
He 50,000 rTC 
H
2
 60,000 rTC 
Calibration method 
Standard gas calibration is fundamentally a substitute gas calibration (see 
page 223) without zero calibration. The correction values are transferred to 
all the analyzer module’s sample components and measurement ranges.