54  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Limas11 IR, Limas21 UV: Preparation for installation 
Installation site 
Ambient temperature 
During operation  with analyzer module installed in a system housing   
+5 to +45 °C  without electronics module 
+5 to +45 °C  with electronics module or only with power supply 
+5 to +40 °C  with electronics module, if I/O modules are installed, or 
only with power supply 
Vibrations/shocks 
max. ±0.04 mm at 5 to 55 Hz; 0.5 g at 55 to 150 Hz 
Sample gas 
Sample gas inlet conditions 
Temperature  The dew point of the sample gas must be at least 5 °C 
lower than the lowest ambient temperature in the overall 
sample gas path. Otherwise, a sample gas cooler or con-
densate trap is required. 
Pressure  The analyzer module is operated under atmospheric pres-
sure; the sample gas outlet is open to atmosphere. Inter-
nal pressure drop < 5 hPa with standard flow rate 60 l/h. 
Permissible absolute pressure range: 800 to 1250 hPa. 
Operation under lower absolute pressure (e.g. at altitudes 
above 2000 m) on request. Overpressure in the sample cell 
max. 500 hPa. 
Flow  20 to 100 l/h 
Remark: Sample gas temperature, pressure and flow rate should be main-
tained constant to such a degree that the fluctuation influence on the accu-
racy of measurement is acceptable (see also chapter "Analyzer module oper-
ating data" (see page 346)).