340  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Uras26: Troubleshooting 
Temperature problem 
Connecting leads of the temperature sensor or the heater discon-
nected 
•  Check the connecting leads and connectors. 
•  Check the fit of the leads in the wire end ferrules. 
Defective thermal link 
•  Check thermal link for continuity and replace if necessary. 
Unstable readings 
Leaks in the gas feed paths 
•  Check the seal integrity (see page 268) of the analyzer module gas paths 
and (if applicable) of the lines to the pneumatics module. 
Vibrations 
•  Provide measures to reduce the vibrations. Permissible vibration levels: 
for analyzer max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz; when in-
stalled in cabinet max. 0.01 ms
−2
 at 0.1 to 200 Hz. 
•  Increase T90; increase switching threshold for non-linear filter (see page 
163).  
Loss of sensitivity 
•  Check the sensitivity variation: 
Indication < 75 %: "Maintenance request" status signal appears. The de-
tector concerned will need to be changed soon. 
Indication < 50 %: The "Failure" status signal appears. 
Uneven emitter modulation 
•  Remove the emitter. 
CAUTION: The emitter temperature is approx. 60 °C in the thermostat 
version of the Uras26!   
•  Check if the chopper wheel turns smoothly. 
•  Check the clamp ring seating. 
•  The chopper wheel should not extend beyond the notch. 
•  Have the emitter and modulator checked by Service.