AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  27
 
 
 
Special requirements for operation 
Special requirements 
•  All cables must enter via the specified cable fittings and be sealed by 
tightening the nuts in accordance with IP54. Cable connectors not in use 
have to be closed with suitable vent plugs. 
•  If the installation site of the gas analyzer is hazardous, 
•  The external connectors on the analyzer module "Power supply 
24 V DC" and "System bus" which are accessible without having to 
open the system housing may not be plugged in or unplugged while 
the power is on. 
•  The system housing may not be opened while the power is on. 
Additional special requirements for operation with flammable sample gas 
•  The sample gas pressure in the gas analyzer must correspond to the 
atmospheric conditions in all operational conditions and in the event of 
faults (positive pressure ≤ 3 hPa). 
•  If pressurized enclosure fails (supply of purge gas to the purge curtain) 
and an alarm is triggered, the fault must be corrected quickly. The gas 
analyzer can remain in operation. If the fault cannot be eliminated 
quickly, supply of the sample gas must be cut off. 
•  If the gas analyzer is not in operation, then the supply gases must be cut 
off. 
•  Tests: 
•  At start-up, after failure of the pressurized enclosure and at regular 
intervals, a test for proper functioning of the pressurized enclosure 
must be conducted by a technically competent person. At this time, 
the conditions stipulated in the section "Technical data" (see page 
25) should be checked and established. 
•  The alarm output must be checked. 
•  The seal integrity of the sample gas path and purge gas path must 
be checked at suitable intervals 
•  If the gas analyzer contains several analyzer modules, the pressurized 
enclosure must be installed as well as monitored and checked separately 
for each individual analyzer module. In the event of a fault, an alarm 
must be triggered in each individual analyzer module. 
NOTES 
The pressurized enclosure (supply of purge gas to the purge curtain) and if 
required, a housing purge must be provided separately.   
The pressurized enclosure is not required if the gas analyzer is used to the 
measure non-flammable gases.