40  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Housing purge 
When does a housing purge become necessary? 
A housing purge becomes necessary when the sample gas contains flamma-
ble, corrosive or toxic components. 
Prerequisite for housing purge 
A housing purge is possible if the system housing is protected to IP54 (with 
connection box) or IP65 (without power supply). The purging gas connectors 
(⅛ NPT female thread) are factory-installed based on orders. 
Caldos25, Caldos27, Magnos206, Magnos28, Magnos27 analyzer modules 
In gas analyzers with the Caldos25, Caldos27, Magnos206, Magnos28 and 
Magnos27 analyzer modules there is a gas-tight separation of the central 
unit and the analyzer. As a result, the central unit and the analyzer can be 
purged both separately (in parallel) and jointly (in series). 
Limas11 IR, Limas21 UV, Limas21 HW, Uras26 and ZO23 analyzer modules 
In gas analyzers with the Limas11 IR, Limas21 UV, Limas21 HW, Uras26 and 
ZO23 analyzer modules there is no gas-tight separation of the central unit 
and the analyzer. They can therefore only be purged together. If separate 
purging of the central unit and the analyzer is required, the analyzer module 
must be installed in a separate system housing protected to IP54. 
Fidas24 and Fidas24 NMHC analyzer modules 
In gas analyzers with the Fidas24 and Fidas24 NMHC analyzer modules the 
housing purge is provided in such a way that a part (approx. 600 to 700 l/h) 
of the instrument air is conducted continuously through the housing as 
purge air. Thus it is made sure that no ignitable gas mixture can form within 
the housing in case of a leak in the combustion gas path. 
NOTE 
The analyzer modules Caldos25, Caldos27, Limas11 IR, Limas21 UV,   
Limas21 HW, Magnos206, Magnos28, Magnos27, Uras26 and ZO23 may not 
be purged in series with the analyzer module Fidas24 and Fidas24 NMHC on 
account of the different requirements for the purge gas supply.   
Central unit with housing purge 
A central unit with no analyzer module installed can also be ordered in the 
version "with housing purge". In this case, the purging gas connectors are 
installed ex works in the end plate. This closes the system housing to the 
rear and at the bottom instead of the analyzer module. 
ATTENTION 
In the case of joint purging of the central unit and analyzer, the purge gas 
must first of all be conducted through the central unit and then through 
the analyzer. If the flow is in the opposite direction and there are leaks in 
the sample gas feed path, corrosive sample gas components can destroy 
the electronic components! 
An analyzer module which is used to measure corrosive sample gas com-
ponents must always be positioned last in the line!