Limas11 IR, Limas21 UV, Limas21 HW, Uras26: Relinearization 
When should relinearization be performed? 
Sample component relinearization should be performed 
•  If the linearity deviation exceeds the permissible value von 1 % of the 
span. 
•  If the start of a suppressed measurement range is to be calibrated. 
•  After a component (lamp/emitter, sample cell, calibration unit/cell, de-
tector) is installed in or removed from a beam path. 
We recommend performing the relinearization of a sample component after 
the measurement range limits have been changed. 
NOTE 
Relinearization is performed separately for each sample component. 
Test gases 
Depending on the number and type of measurement ranges, test gases with 
varying concentrations are required for relinearization: 
Number and type of meas. ranges  Test gas concentration 
1 measurement range  Approx. 40 to 60 % measurement 
range end value ("center point gas") 
2 measurement ranges  End value of the smaller measure-
ment range 
2 measurement ranges,   
1 of which is suppressed. 
Start value of the suppressed 
measurement range 
Menu path 
MENU → Maintenance/Test → Analyzer spec. adjustm. → 
Relinearization 
Relinearization procedure 
1  Perform (see page 274) the basic calibration for the zero and span of the 
sample component to be relinearized. 
2  Select the Relinearization menu item. 
3  Select the Sample component. 
4  Turn on the test gas supply. 
5  Enter the test gas concentration set point value. 
6  Wait until the measurement value reading stabilizes and activate the 
alignment procedure. 
7  Repeat steps 3 to 6 for all sample components.