AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  57
 
 
 
Limas21 HW: Preparation for installation 
Installation site 
Ambient temperature 
+15 to +35 °C with installation in a system housing with or without an elec-
tronics module 
Vibrations/shocks 
max. ±0.04 mm / 0.5 g at 5 to 150 Hz 
Sample gas 
Sample gas composition 
Stack gas from incinerators without sulfur content, SO
2
 concentration 
< 25 ppm, H
2
O concentration < 20 vol.%, filtered with pore size ≤ 0.5 µm 
Sample gas inlet conditions 
Temperature  Sample gas dew point ≤ 65 °C 
Pressure    The analyzer module is operated under atmospheric pres-
sure; the sample gas outlet is open to atmosphere. Inter-
nal pressure drop < 5 hPa with standard flow rate 60 l/h. 
Permissible absolute pressure range: 800 to 1250 hPa. 
Operation under lower absolute pressure (e.g. at altitudes 
above 2000 m) on request. Overpressure in the sample cell 
max. 500 hPa. 
Flow  20 to 90 l/h 
Remark: Sample gas temperature, pressure and flow rate should be main-
tained constant to such a degree that the fluctuation influence on the accu-
racy of measurement is acceptable (see also chapter "Analyzer module oper-
ating data" (see page 346)). 
Test gases 
Zero-point calibration 
Nitrogen or air or UV sample component-free gas 
End-point calibration 
Calibration cells or test gas for each sample component 
Dew point 
The dew point of the test gases must be approximately the same as the dew 
point of the sample gas. 
NOTE 
The Notes for calibrating (see page 249) must be considered. 
Pressure sensor 
The pressure sensor is installed in the gas analyzer ex works.   
The pressure sensor is located in the sample gas feed path if the internal gas 
lines are executed as FPM flexible tubes. The connection of the pressure 
sensor is led to the outside via an FPM hose if the internal gas lines are exe-
cuted as tubes. The connection of the pressure sensor is documented in the 
pneumatic diagram contained in the analyzer data sheet.