AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  63
 
 
 
Magnos27: Preparation for installation 
Installation site 
Ambient temperature 
During operation  with analyzer module installed in a system housing   
+5 to +45 °C  without electronics module 
+5 to +45 °C  with electronics module or with power supply only 
+5 to +50 °C  without electronics module or Uras26 and version with 
sample chamber direct connection 
Vibrations/shocks 
max. ±0.04 mm at 5 to 60 Hz 
Sample gas 
Sample gas inlet conditions 
Temperature  The dew point of the sample gas must be at least 5 °C 
lower than the lowest ambient temperature in the overall 
sample gas path. Otherwise, a sample gas cooler or con-
densate trap is required.   
When there is a direct sample chamber connection the 
maximum sample gas dew point is 55 °C.   
Fluctuations in water vapor content cause volume errors. 
Pressure    The analyzer module is operated under atmospheric pres-
sure; the sample gas outlet is open to atmosphere. Inter-
nal pressure drop < 5 hPa with standard flow rate 60 l/h. 
Permissible absolute pressure range: 800 to 1250 hPa. 
Operation under lower absolute pressure (e.g. at altitudes 
above 2000 m) on request. Overpressure in the sample 
chamber max. 100 hPa. 
Flow  20 to 90 l/h 
Remark: Sample gas temperature, pressure and flow rate should be main-
tained constant to such a degree that the fluctuation influence on the accu-
racy of measurement is acceptable (see also chapter "Analyzer module oper-
ating data" (see page 346)). 
Flammable gases 
The gas analyzer may not be used for the measurement of flammable gases. 
Test gases 
Zero-point calibration 
Oxygen-free process gas or Nitrogen 
End-point calibration   
Process gas with a known oxygen concentration or a substitute gas, e.g. 
dried air   
Dew point 
The dew point of the test gases must be approximately the same as the dew 
point of the sample gas. 
NOTE 
The notes for calibrating (see page 259) must be considered.