48  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Fidas24: Preparation for installation 
Installation site 
Ambient temperature 
During operation  with analyzer module installed in a system housing 
+5 to +45 °C  without electronics module 
+5 to +45 °C  with electronics module or with power supply only 
Vibrations/shocks 
max. 0.5 g, max. 150 Hz 
Sample gas 
Sample components 
Hydrocarbons. The concentration of the gas components in the sample gas 
path must not exceed the temperature-dependent LEL. The analyzer tem-
perature is 180 °C. 
Sample gas inlet conditions 
Temperature  ≤ Thermostat temperature (thermostat temperature for 
sample gas path, detector and air jet injector ≤ 200 °C, 
pre-set ex works to 180 °C) 
Inlet pressure  p
abs
 = 800 to 1100 hPa 
Flow  Approx. 80 to 100 l/h at atmospheric pressure (1000 hPa) 
Humidity  ≤ 40 % H
2
O 
Remark: Sample gas temperature, pressure and flow rate should be main-
tained constant to such a degree that the fluctuation influence on the accu-
racy of measurement is acceptable (see also chapter "Analyzer module oper-
ating data" (see page 346)). 
Sample gas outlet conditions 
The outlet pressure must be the same as the atmospheric pressure. 
Flammable gases 
The gas analyzer may be used for the measurement of flammable gases 
provided that the flammable portion does not exceed 15 vol.% CH
4
 or 
C1 equivalents. 
Further sample gas inlet conditions 
The sample gas may not be explosive at any time. 
The analyzer module must not be used for measurement of gases containing 
organometallic compounds, e.g. leaded gasoline additives or silicone oils.