AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  65
 
 
 
Uras26: Preparation for installation 
Installation site 
Ambient temperature 
During operation  with analyzer module installed in a system housing   
+5 to +45 °C  without electronics module 
+5 to +40 °C  with electronics module or with power supply only 
Vibrations/shocks 
max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz; slight transient effect on 
measured value in the region of the beam modulation frequency 
Sample gas 
Sample gas inlet conditions 
Temperature  The dew point of the sample gas must be at least 5 °C 
lower than the lowest ambient temperature in the overall 
sample gas path. Otherwise, a sample gas cooler or con-
densate trap is required. 
Pressure    The analyzer module is operated under atmospheric pres-
sure; the sample gas outlet is open to atmosphere. Inter-
nal pressure drop < 5 hPa with standard flow rate 60 l/h. 
Permissible absolute pressure range: 800 to 1250 hPa. 
Operation under lower absolute pressure (e.g. at altitudes 
above 2000 m) on request. Overpressure in the sample cell 
max. 500 hPa. 
Flow  20 to 100 l/h 
Remark: Sample gas temperature, pressure and flow rate should be main-
tained constant to such a degree that the fluctuation influence on the accu-
racy of measurement is acceptable (see also chapter "Analyzer module oper-
ating data" (see page 346)). 
Corrosive gases 
Highly corrosive associated gas components such as chlorine (Cl
2
 or hydro-
gen chlorides (e.g. wet HCl) as well as gases or aerosols containing chlorine 
must be cooled or pre-absorbed. A housing purge must be provided. 
Flammable gases 
The analyzer module is suitable for the measurement of flammable gases 
and vapors under atmospheric conditions (p
abs
 ≤ 1.1 bar, oxygen content 
≤ 21 vol.%). Temperature class: T4. 
The sample gas may not be potentially explosive in normal operation; if it is 
potentially explosive in the event of faults in the sample gas supply, then 
only rarely and for a short time (corresponding to Zone 2). 
Pressure in the sample gas feed path in normal operation p
e
 ≤ 100 hPa; in the 
event of faults in the sample gas supply, the pressure may not exceed the 
maximum value of p
e
 = 500 hPa. 
For the measurement of flammable gases and vapors, the version with in-
ternal gas lines made of stainless steel should be selected and a housing 
purge with nitrogen provided. 
Prior to using the analyzer module, the corrosion resistance of the sample 
gas must be tested.