AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  127
 
 
 
Gas path and housing initial purge 
Purge before commissioning 
The gas paths and if necessary, the system housing should be purged prior 
to starting the gas analyzer. 
Firstly this ensures that the gas paths and system housing are free from 
contaminants e.g. corrosive gases, and accumulations of dust on start-up. 
Secondly this prevents any explosive gas/air mixture present in the gas 
paths or system housing being ignited when the power supply is connected. 
Purge gas 
Nitrogen or instrument air should be used as a purge gas. 
Purge gas flow rate during initial purge 
The purge gas flow and the duration of the purge process depend on the 
volume to be purged (see the following table). If the purge gas flow is lower 
than specified, the duration of the purging must be extended accordingly. 
Volume to be purged  Purge gas flow rate  Duration 
Gas path  100 l/h (max.)  approx. 20 s 
Central unit with or without analyzer 
module 
200 l/h (max.)  approx. 1 h 
Analyzer separately: Caldos25, Caldos27, 
Magnos206, Magnos28, Magnos27 
200 l/h (max.)  approx. 3 min 
NOTE 
The purge gas flow rates given in the table only apply to the initial purge. 
Other values apply during operation (see page 136). 
 
ATTENTION 
Purge gas can escape from the housing if there are any leak points. 
When using nitrogen as the purge gas, take all required precautions 
against suffocation.   
The purge gas flow must always be restricted upstream of the purge gas 
inlet! If the purge gas flow is not restricted until after the purge gas out-
let, the full pressure of the purge gas will affect the housing seals which 
can destroy the operator panel keypad!