216  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Automatic calibration 
Definition 
Automatic calibration means: Zero and span calibration run automatically 
after starting. 
Test gas supply 
The test gas supply can be started automatically by means of the pneumatic 
module’s solenoid valves or via external solenoid valves. 
Depending on the gas supply circuit and the number of analyzer modules 
installed there are several layout possibilities for the test gas supply (see 
page 219).   
Dew point 
The dew point of the test gases must be approximately the same as the dew 
point of the sample gas. 
Analyzer modules with several detectors 
In analyzer modules with several detectors (e.g. Uras26) all detectors are 
calibrated simultaneously. 
Starting the automatic calibration 
Automatic calibration is started 
•  at time intervals determined by the internal clock   
•  by an external control signal or 
•  manually via the gas analyzer’s display and control unit. 
Internal start 
Automatic calibration is normally started cyclically on a time-controlled ba-
sis by the internal clock of the gas analyzer. 
The cycle time is provided with a parameter with the calibration data (see 
page 227). 
External start 
The "Start automatic calibration" control signal is needed for external start-
ing of automatic calibration: 
Level  Low 0 to 3 V → High 12 to 24 V edge. The Low → High transition 
may also be generated by an external contact. After the transi-
tion the High level must be present for at least 1 s. 
Input  Digital input DI1 on Digital I/O module – "Status signals/ Exter-
nally controlled calibration" (see page 112) (standard function 
block application)