AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  255
 
 
 
Magnos28: Notes for calibrating 
Sample components 
The Magnos28 analyzer module has at least one sample component normally 
with 4 measuring ranges. 
Test gases 
Zero calibration: Oxygen-free process gas or substitute gas 
Span calibration: Process gas with a known oxygen concentration or sub-
stitute gas, e.g. dried air 
Highly suppressed measuring ranges (≥95 to 100 vol.% of O
2
) should only be 
calibrated with test gases with O
2 
concentration in the selected measure-
ment range. 
Test gas for corrected sample components 
During calibration possible electronic cross-sensitivity and/or carrier gas 
corrections by other measurement components are switched off. Therefore, 
corrected measurement components should be calibrated only using a test 
gas consisting of the measurement component and an inert gas like Nitro-
gen. 
Calibration with substitute gas 
If test gases are not available for calibration, the analyzer module can be 
factory-set for calibration with a substitute gas (see also section "Calibra-
tion methods" (see page 223)). This setting is documented in the analyzer 
data sheet. 
Substitute gas calibration of the Magnos28 analyzer module is described 
using the example of "Purity measurement of CO
2
" (see page 258). 
Single-point calibration 
Single-point calibration of the Magnos28 analyzer module is described in the 
"Magnos28: Single-point calibration" section (see page 256). 
Suppressed measuring ranges 
If suppressed measurement ranges with a suppression ratio of ≥ 1:5 are set 
in the Magnos28 analyzer module, the pressure sensor has been specially 
adjusted at the factory. Therefore, only a common calibration should be 
carried out (not a single or substitute gas calibration) in this case. 
Sequence of calibration 
Zero-point calibration must always precede end-point calibration.   
Wait until the warm-up phase has ended 
The analyzer module should only be calibrated after the warm-up phase (see 
page 135).