AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  347
 
 
 
Caldos27: Operating specifications 
The stability data apply only to measurement ranges ≥ Class 2. 
 
Linearity deviation  ≤ 2 % of span 
Repeatability  ≤ 1 % of span 
Zero drift  ≤ 2 % of the smallest possible measurement range per week 
Sensitivity drift  ≤ 0.5 % of the smallest possible measurement range per week 
Output fluctuation (2 
σ) 
≤ 0.5 % of smallest measurement range span at electronic T90 time = 0 s 
Detection limit (4 
σ) 
≤ 1 % of smallest measurement range span at electronic T90 time = 0 s 
Flow influence  ≤ 0.5 % of span at a flow change of ±10 l/h.   
At the same sample and test gas flow rate, the flow influence is automatical-
ly compensated. 
Associated gas effect  The knowledge of the sample gas composition is necessary for the analyzer 
configuration. 
Temperature influence  At each point in the measurement range: ≤ 0.5 % of span per 10 °C, with re-
spect to the temperature during calibration. Thermostat temperature = 
60 °C; Ambient temperature in permissible range 
Air pressure effect  ≤ 0.25 % of span per 10 hPa for the smallest possible ranges given; in larger 
measurement ranges the influence effect is correspondingly smaller.   
Working range of installed pressure sensor: p
abs
 = 600 to 1250 hPa   
Option: Operating altitude over 2000 m 
Power supply influence  24 V DC ± 5 %: ≤ 0.2 % of span 
Position influence  < 1 % of span up to 30° deviation from horizontal orientation 
T
90
 time  T
90
 ≤ 2 s with direct sample chamber connection and sample gas flow = 
60 l/h (applies to a gas analyzer with one analyzer module)