42  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Power supply 
Gas analyzer power supply 
A power supply is installed in the central unit of the gas analyzer. It provides 
the supply voltage to power the electronics module. 
Analyzer module power supply 
The analyzer module requires a 24 V DC voltage supply ± 5 %. 
If the analyzer module is installed in the central unit, power can be supplied 
by the central unit power supply. 
If the analyzer module is installed in a separate system housing rather than 
in the central unit, a distinction has to be made between three housings: 
•  The analyzer module can be powered by the central unit power supply if 
the optional power line filter -Z01 is installed in the central unit and no 
analyzer module is installed in the central unit. 
•  If only one analyzer module is installed in the system housing, an 
AO2000 power supply can be installed in the same system housing. 
•  If two analyzer modules are installed in the (separate) system housing, a 
power supply outside the system housing must be provided. This power 
supply unit must comply with the specifications of the AO2000 power 
supply unit. 
NOTE 
Only 1 analyzer module may be supplied with 24 VDC from the central unit 
power supply! A separate 24 VDC supply is required for additional analyzer 
modules. 
Power supply 
The central unit power supply provides 24 VDC for the electronics module 
and one analyzer module installed in the central unit or one external analyzer 
module. 
Input voltage  100 to 240 VAC, −15 %, +10 % 
Input current  max. 2.2 A 
Line frequency range  50 to 60 Hz ± 3 Hz 
Power consumption  max. 187 VA 
Output voltage  24 V DC ± 3 % 
Connection 3-pin grounded-instrument connector to 
EN 60320-1/C14, connecting cable supplied