322  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Status messages 
List layout 
The status message list contains the following information: 
No.  The status message number appears in the menu line display 
Text  Full text of the status message is shown in the detailed display 
S  x = Status message sets the overall status 
A  x = Status message sets the "Error" individual status 
W  x = Status message sets the "Maintenance request" individual 
status 
F  x = Status message sets the "Maintenance mode" individual 
status 
Reaction/ 
Remark 
Explanations and corrective measures in case of status mes-
sages 
Status messages 
No. Text  S A 
F Reaction/Remark 
Run time error             
1– 
21 
Runtime error 1 to   
Runtime error 21 
        Notify Service if these messages occur repeatedly. 
System controller             
101  System controller shut down at                  For information; shows date and time 
102  System controller starts up at                  For information; shows date and time as well as warm or 
cold start 
103  Install Module:          For information 
104  Remove Module:          For information 
105  Reactivate Module:          For information 
106  A user installed module:          For information 
107  A user removed module:          For information 
108  A user replaced module:          For information 
109  A password is active! To delete, 
press the <MEAS> key on the 
measurement value display. 
                For information on Password Protection see "Password 
protection" (see page 149) section; not logged   
110  System booting.          Not logged 
111  This system is currently under 
remote control 
        Not logged 
112  Display/control unit synchro-
nizing with analyzer. Please 
wait. 
        Not logged 
113  The system time was changed 
from -> to: 
        Not logged 
114  The system is saving the 
changed parameters. Please 
wait.