158  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Changing the number of decimal places 
Menu path 
MENU → Configure → Component-Specific → Measurement Range → 
Select Component → ... 
Selection 
All measurement ranges configured (at the factory) for a sample component 
are displayed. 
Procedure 
Select the measurement range with the arrow keys, press SET DECIMAL, set 
the number of decimal places with the arrow keys and press ENTER to con-
firm. 
NOTES 
The setting only affects the values indicated on the screen. 
The number of decimal places cannot be increased. 
The altered number of decimal places is shown on the screen after switching 
to measurement mode. 
Number of decimal places 
When the screen displays the measured value in physical units (e.g. ppm) the 
number of places after the decimal point depends on the size of the set 
measurement range: 
Span  Places after the decimal point 
≤ 0.05  5 
≤ 0.5  4 
≤ 5  3 
≤ 50  2 
≤ 500  1 
> 500  0 
For the display of the measured value as a percentage of the whole measur-
ing range (%Span) two places are always shown after the decimal point. 
When setting the parameters, the number of decimal places is the same as 
in the display in measuring mode.