282  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
CAUTION! 
The heated sample gas connection is hot (approx. 180 °C)! Allow the sam-
ple gas connection to cool down after switching off the power supply 
(approx. 30 minutes). 
1  Turn off the sample gas supply to the analyzer module.   
Switch off 115/230-V-AC power supply of the gas analyzer and heater as 
well as if necessary, the separate 24-V-DC supply of the analyzer module. 
2  Unscrew exhaust pipe 2 from the exhaust air outlet 1 (key size 12 mm). 
3  Unscrew air jet injector from the exhaust air outlet 1 (key size 14 mm). 
4  Clean air jet injector in ultrasonic bath. Use aqueous cleaning agent (e.g. 
Extran). 
5  Replace O-rings 3, 4 and 5 with new O-rings. 
NOTE: When cleaning the air jet injector always replace all three O-rings. 
Contaminated or damaged O-rings impair the suction performance of 
the air jet injector; this may result in the failure of the Fidas24. 
Slightly lubricate all three O-rings with high temperature grease (part 
no. 772341) before mounting them. 
6  Screw air jet injector 6 into the exhaust outlet. Make sure that the 
O-rings are seated correctly. 
7  Screw the exhaust pipe 2 onto the 1 exhaust air outlet. 
8  Reconnect the sample gas supply to the analyzer module. 
9  Switch on the power supply. 
10  Check output variables of the internal pressure regulator for the supply 
gases and adjust if necessary (see section "Fidas24: Gas analyzer 
start-up" (see page 129)).   
11  Calibrate the gas analyzer at the end of the warm-up phase.