AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  331
 
 
 
No. Text  S A 
F Reaction/Remark 
Miscellaneous Messages             
1000  This function block has an er-
ror: 
x  x      Is supplemented with a reference to the function block 
type. 
1001 Condensate penetration.          Overall message for guiding reaction to condensate 
penetration; not logged 
1002  Flow rate too high at this point!  x  x          Currently not used 
1003  Flow rate too low at this point!  x  x      Currently not used 
System Cooler             
1100  Cooler temperature too high.  x  x          Sample gas feed module pump is automatically turned 
off. Check the system cooler and sample gas prepara-
tion system. 
1101  Cooler temperature too low.  x  x         
1102  Condensate penetration in 
cooler.  
x  x     
1103  Flow rate too low in cooler.          x      Check the system cooler and sample gas preparation 
system. 
1104  Cooler condensate level too 
high. 
    x     
1105  Cooler condensate level too 
high. 
    x    Empty the condensate bottle. 
1106  Cooler reagent level too low.          x      Fill the reagent container.