General Technical Data
Delivery Your Accu-Chek Spirit Combo insulin pump delivers insulin according to the programmed
hourly basal rate, 24 hours a day.
j Minimal dose: 1/20 of the hourly basal rate, maximal 20 times per hour.
j Test interval: every 3 minutes.
Basal rate Min. 0.05 U / h, max. 25 U / h. There are 24 hourly basal rates adjustable in 0.01
(up to 1.00 U / h), 0.05 (up to 10.0 U / h) and 0.1 (up to 25.0 U / h) unit increments.
Bolus The max. bolus amount per delivery is 25 insulin units. The bolus amount for the Quick
Bolus is adjustable in increments of 0.1, 0.2, 0.5, 1.0 and 2.0 units. For the Standard Bolus,
Extended Bolus and Multiwave Bolus, the amount is adjustable in xed increments of
0.1 units. The duration of the Extended Bolus and the Multiwave Bolus is adjustable in
intervals of 15 minutes (15 minutes up to 12 hours).
Temporary Basal Rate Adjustable in 10 % increments, 0 – 90 % for decreases, 110 – 250 % for increases. The
duration is adjustable in 15 minute intervals, up to a maximum of 24 hours. The last
programmed duration is given by default for the next Temporary Basal Rate change
programmed.