General Technical Data
Maximum time for an Error
E4: OCCLUSION**
Plastic cartridges:
j at medium basal rate 1.0 U / h: ≤ 2 h
j at minimum basal rate 0.05 U / h: ≤ 24 h
Maximum volume before Error
E4: OCCLUSION**
Plastic cartridges: ≤ 1.5 U
Maximum pressure 400 kPa (4.0 bar), for plastic cartridges and infusion sets.
Flow (delivery rate) During infusion set priming and bolus: 0.2 U / sec.
Maximum quantity delivered at
a single fault condition
≤ 1.0 U
Cartridge Accu-Chek Spirit 3.15 ml Cartridge System and Accu-Chek 3.15 ml Plastic Cartridge with a
luer connection.