GB 04270 - Edition 02 - March 12
3/4
C 404
Maintenance Menu
Events history
RS 232 Code Description
History export at .CSV format
Date and Time of
the event
Event = Error (Exxx)
= Warning (Wxxx)
D 100
= Event (Ixxx)
Events history records the last 30 events appeared. Beyond 30, the
oldest recorded event will be replaced by the last one appeared and
so on.
Events list
DescriptionEventCode
Air inletAir inletI300
Automatic cycle stop if pollution (He signal measured > He max)
Stp Cy He>Hemax
I301
Primary pump counter resetRst count RVP
I302
High vacuum pump 1 counter resetRst count TMP1
I303
High vacuum pump 2 counter resetRst count TMP2
I304
High vacuum pump 3 counter resetRst count TMP3
I305
Reset filament 1 counterRst count Fil1
I306
Reset filament 2 counterRst count Fil2
I307
Reset cycles counterRst count cycle
I308
Automatic Ie increase to 1,5 mA for He/3He or 0,6 mA for HyIe increase
I309
Automatic autocalibration restartAutocal restart
I310
Date or time changeDate/Time updat
I313
Complete detector parameters resetFull param rst
I318
Filament change (manually or automatically with firmware)Fil change
I319
Automatic internal Pirani gauge calibrationCalib. Pirani int
I320
Storage delay : it is activated if the leak detector doesn't start since 15 days
(minimum).
Storage delay
I321
adixen Vacuum Products - ASM 310 Operating instructions