Calibration checking Autocalibration ON, the function "calibration checking" does a
calibration checking according to set parameters.
The calibration checking is done with the internal calibrated leak
(location parameter = internal).
The calibration checking function allows to the operator to gain
time because the calibration checking is faster than the complete
calibration.
The leak detector compares the internal calibrated leak delivered
signal to the calibrated leak parameters memorized:
If the value is correct, the leak detector calibration is good.
If the value is out of limits, there is a message which request to
start acomplete leak detector calibration.
Function activation.
Set points launching the
calibration checking. The
first reached will launch the
checking.
•Atanytime,theoperatorcanlaunchaleakdetectorcalibration
checking:detectorinstand-bymode,presstwiceonthefunction
key
inlessthan5secondes
Other advanced methods
It is also possible to do a detector autocalibration with an external
calibrated leak
D 410.
•Itisadvisedtoperformanautocalibration:
-atthedetectorswitchingoninordertomakesurethatitisin
correctconditionsuse
-tooptimizethemeasurereliabilityforhighsensitivitytests
-incaseofdoubtaboutthegoodleakdetectorfunctioning(ability
todetectcorrectlyaHeleak).Atanytime,aninternalcalibration
canbelaunched.
-incaseofcontinuousandintensiveuse:launchaninternalcalibra-
tionateachteambeginning(8hoursofwork).