4/15
C 406
Advanced Menu
GB 04708 - Edition 03 - March 12
Calibration checking Autocalibration ON, the function "calibration checking" does a
calibration checking according to set parameters.
The calibration checking is done with the internal calibrated leak
(location parameter = internal).
The calibration checking function allows to the operator to gain
time because the calibration checking is faster than the complete
calibration.
The leak detector compares the internal calibrated leak delivered
signal to the calibrated leak parameters memorized:
If the value is correct, the leak detector calibration is good.
If the value is out of limits, there is a message which request to
start acomplete leak detector calibration.
Function activation.
Set points launching the
calibration checking. The
first reached will launch the
checking.
•Atanytime,theoperatorcanlaunchaleakdetectorcalibration
checking:detectorinstand-bymode,presstwiceonthefunction
key
inlessthan5secondes
Other advanced methods
It is also possible to do a detector autocalibration with an external
calibrated leak
D 410.
•Itisadvisedtoperformanautocalibration:
-atthedetectorswitchingoninordertomakesurethatitisin
correctconditionsuse
-tooptimizethemeasurereliabilityforhighsensitivitytests
-incaseofdoubtaboutthegoodleakdetectorfunctioning(ability
todetectcorrectlyaHeleak).Atanytime,aninternalcalibration
canbelaunched.
-incaseofcontinuousandintensiveuse:launchaninternalcalibra-
tionateachteambeginning(8hoursofwork).
adixen Vacuum Products - ASM 310 Operating instructions