GB 03295 - Edition 01 - June 07
1/1
C 110
Perform a test
By default, the leak detector is set to perform a helium hard vacuum
test, in the most sensitive test mode.
Hard vacuum test mode Sniffing test mode
Select the hard vacuum
test mode (
C 402
).
Select the sniffing
test mode (
C 402
).
Connect the part to be tested to the leak
detector inlet port or put the part in the test
chamber connected to the leak detector.
While the leak detector is in stand-by, connect
the sniffer probe (accessory to be purchased
separately) to the sniffer port of the leak
detector.
Set the reject point (
C 401
).
START
STAND-BY
Start a test.
The leak value measured displays.
START
STAND-BY
Stop the test.
The detector returns automatically to stand-by mode
.
If Memo function is activated and a printer is connected to the detector, a test ticket
will be printed automatically after the test (
C 406).
2 tests modes are proposed: hard vacuum and sniffing (
A 200).
Notes
In stand-by mode, the leak value displayed is the helium
background value.
If the automatic cycle function is activated, the test end is
different
(
C 304).
adixen Vacuum Products - ASM 310 Operating instructions