Adixen ASM 310 - Test Mode

Adixen ASM 310
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2/5
C 402
Test Menu
GB 03341 - Edition 03 - March 2010
A hard vacuum test can be performed as soon as one of the test
modes is reached
The leak detector will automatically switch to the selected mode as
soon as the inlet pressure has crossed the threshold.
Test mode
Gross leak mode
crossing set point
Normal mode
crossing set point
High Sensitivity mode
crossing set point
Pressure
Roughing
Gross Leak mode
Normal mode
High Sensitivity mode
Use the correction factor to work in a unit other one than those
proposed (See board
G 700 for the factor to be applied) or with
an installation having its own pumping.
According to the He concentration level in the gas used for the
leak detection, the displayed signal changes.
Example: signal displayed with a calibrated leak of 1x10
-7
mbar.l/s
(with 100 % He) connected to the detector inlet
% He in the
used gas
100
% 50 % 10 % 1 %
Displayed signal
on the leak detector
without COR
1 x 1
0
-7
mbar.l/s
5 x 1
0
-8
mbar.l/s
1
x 10
-8
mbar.l/s
1
x 10
-9
mbar.l/s
COR value 1 2 10 1
00
Displayed signal
on the leak
detector with COR
1 x 10
-7
mbar.l/s
Leak detector calibration with a pumping group E 411.
By default, the leak detector is configurated to work in vacuum
test, in the most sensitive test mode: this configuration meet the
majority of the operators needs.