Adixen ASM 310 - Use and Display (Ctd); Autocalibration (Remote Control); Inlet Vent;Sniffing Test Mode; External Calibration

Adixen ASM 310
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GB 03337 - Edition 02 - July 2011
3/3
C 700
Remote control (accessory)
Inlet vent
Sniffing test mode
VENT SNIF
For these 2 functions, the remote control displays only their status:
use
the control panel in order to activate/deactivate them.
C 402
Display
The LED indicator is ON/OFF when the function is activated/
deactivated.
External calibration The remote control displays only its status: use the control panel in
order to start/stop it.
Display
The LED indicator is blinking when the external calibration is
activated.
At the end of the calibration procedure, the LED indicator is OFF
and the correction factor COR is ON.
External calibration
E 410
mbar.l/s
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
10
-3
10
-2
Autocalibration
In order to start an internal auto-calibration, the operator can
use the AUTOCAL function key on either the control panel or the
remote control.
Display
The LED indicator is ON when the control key is activated.
The LED indicator is OFF at the end of the calibration procedure.
Calibration of the leak detector
C 406
In order to start the zero function, operator can use the ZERO
control key on either the control panel or the remote control.
Display
Two indicators are ON when the zero function is activated:
the ZERO indicator,
the zero scale indicator.
Zero function
C 406
Use and display
(ctd)
Zero function
adixen Vacuum Products - ASM 310 Operating instructions