GB 04708 - Edition 03 - March 12
5/15
C 406
Advanced Menu
Leak detection:
Analyzer cell
Parameters available in this menu are mainly used for a leak
detector manual calibration.
THIS TYPE OF CALIBRATION IS KEPT FOR SERVICE CENTERS.
Filament switching on.
Filament selection.
Parameters for manual
calibration.
There are 2 filaments in the analyzer cell. The operator can select
the one or the other one.
The filament can be switched off by the operator (OFF): it is not
necessary to switch it off in stand-by mode to save it.
Filament
•Donotswitchoffthelamentexcepttodoamanualcalibration.
•Attention:themanualcalibrationiskepttoleakdetection
experts.
•Theleakdetectorswingsautomaticallyfromafilamenttothe
otheroneiftheselectedfilamentinprogressofusebecomes
defective.
•Attheleakdetectorswitchingon,thedetectorwillstartagainwith
theselectedlamentatthelastswitchingoff.
adixen Vacuum Products - ASM 310 Operating instructions