GB 03316 - Edition 02 - January 09
C 401
Set points Menu
3/3
The purpose of this threshold is to check that the sniffer probe is
operational.
When the helium signal is lower than the set "probe clogged" reject
point, an information will be communicated to the operator to
check the probe.
Sniffer probe maintenance
G 400
Probe clogged
Block the sniffer probe end from time to time with a finger to
check that the helium signal goes down. If not, the probe may be
clogged.
Do not block the end a long time: if the measured signal decreases
too much, there is risk of taking out of the sniffing test.
•
•
adixen Vacuum Products - ASM 310 Operating instructions