GB 03341 - Edition 03 - March 2010
3/5
C 402
Test Menu
Setting
1
= Auto ;
2
= off ;
3
= 10 sec
T = 0 .....................: Cycle start, roughing.
T = x s ..................: Test pressure reached, measure time counter
start (x
< 10 s, otherwise part bad).
T = x + 10 sec.......: Test stop, stand-by mode, inlet vent if
automatic
If "automatic", set it
Compulsory checking: test time.
When reached, measured leak
value display.
Example
Activated function if "Automatic"
Optional checking: roughing time
maxi. If crossed = part bad.
Function to use to automate small production or realize repetitive
operations with different detectors.
•
1
2
3
Optional activation:
roughing time checked
This function allows an automatic control of the roughing time and
measurement time in hard vacuum test.
Cycle end function
adixen Vacuum Products - ASM 310 Operating instructions