EasyManua.ls Logo

Adixen ASM 310 - He Calibrated Leaks

Adixen ASM 310
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2/16
A 400
Accessories
GB 03329 - Edition 06 - February 12
C1
He calibrated leaks
(from 10
-9
mbar l/s to 10
-4
mbar l/s)
Decade
Range
code
Standard value provided Reservoir
10
-9
mbar l/s 9 between 1 and 3 · 10
-9
mbar l/s Aluminium
10
-8
mbar l/s 8 between 1 and 3 · 10
-8
mbar l/s Aluminium
10
-7
mbar l/s
7 between 1 and 3 · 10
-7
mbar l/s Aluminium
3 between 3 and 6 · 10
-7
mbar l/s 250 cc stainless steel
10
-6
mbar l/s
6 between 1 and 3 · 10
-6
mbar l/s 250 cc stainless steel
5 between 5 and 8 · 10
-6
mbar l/s 250 cc stainless steel
10
-5
mbar l/s
4 between 1 and 3 · 10
-5
mbar l/s 1000 cc stainless steel
2 between 5 and 8 · 10
-5
mbar l/s 1000 cc stainless steel
10
-4
mbar l/s 1 between 1 and 3 · 10
-4
mbar l/s 1000 cc stainless steel
Supplied with certificate of calibration.
± 10 % accuracy.
Possibility to deliver a standard value (included in a
predeterminated range, see table below) or a specific value
according to the customer requirement.
adixen Vacuum Products - ASM 310 Operating instructions