GB 04708 - Edition 03 - March 12
9/15
C 406
Advanced Menu
B
Calibration checking with
an internal leak
CALIBRATION INFORMATIONS:
DATE|Jan/08/2007 TIME|01:37:20
current internal temperature(C)| 32
current coef.sens| 00.54
global rate| 1.91E-07
background rate| 6.39E-11
calibrated leak-rate| 1.191E-07
target value| 1.89E-07
percent allowance (+/-)| 15
RESULT(%): 1
C
Test ticket
DATE:Jan/08/2007
HOUR CASE PRESSURE LEAKRATE
00:28:26 start 3.8E+01 3.4E-11
00:28:55 GL 1.2E-02 -------
00:29:40 NR 6.3E-05 -------
00:29:45 stop 5.8E-05 1.1E-05
NR
DATE:Jan/08/2007
HOUR CASE PRESSURE LEAKRATE
01:02:31 start 6.5E+01 4.8E-11
01:02:32 GL 4.6E+01 -------
01:02:33 NR 5.9E-02 -------
01:02:36 stop 9.9E-05 1.3E-05
NR FAIL
Leak value
00:29:45 stop 5.8E-05 1.1E-05
Leak value
01:02:36 stop 9.9E-05 1.3E-05
Test result if Memo
function activated
01:02:36 stop 9.9E-05 1.3E-05
Memo function C 406
adixen Vacuum Products - ASM 310 Operating instructions