2/3
C 700
Remote control (accessory)
GB 03337 - Edition 02 - July 2011
3 different units can be selected in the leak detector but only one
unit is available on the remote control.
So the operator should choose the remote control according to of
the operation unit chosen.
Units C 405
Accessories
A 400
When the operator connects the remote control to the leak detector,
the leak detector unit is automatically reprogrammed with the unit
of the remote control. The remote control unit is memorized by the
detector when the operator disconnects the remote control
.
The remote control:
displays the leak rate value,
controls the start/stop cycle, zero function and internal calibration,
displays the status of the air inlet vent, sniffing test mode and
external calibration,
does not allow to adjust leak detector parameters.
On the remote control and control panel, the displayed values on
the analog and digital displays are exactly the same.
In order to start/stop a cycle, operator can use either the
START
STAND-BY
control key on the control panel or remote control.
Display
The LED indicator is ON/OFF when the control key is activated/
deactivated: if the led is OFF, the leak detector is in stand-by.
As soon as the detector is in test, the TEST indicator is ON.
Connecting
Choice
Analog and digital displays
Use and display
To start/stop a cycle
B 210
START
STAND-BY
adixen Vacuum Products - ASM 310 Operating instructions