Leak detection:
Calibration
It allows to check that the detector is adjusted to detect the selected
tracer gas and display a correct leak value.
Generally, to calibrate the leak detector, a calibrated leak is used as
reference.
Our leak detectors are equipped with an internal He calibrated leak
with reservoir and compensation sensor in temperature.
Setting
Autocalibration activation.
Calibration checking
function.
By default, autocalibration is set operator and the internal calibrated
leak is selected to allow a quick leak detector autocalibration.
Calibration = operator
Calibration must be start by the operator (
key) at the
leak detector switch on.
If calibration is not realized in the next 20 mn after the leak
detector switch on, a reminder message displays.
Calibration = starting
Calibration is automatically start at the leak detector switch on.
No
calibration
Calibration
start