C 406
Advanced Menu
GB 04708 - Edition 03 - March 12
This function freezes the screen with the last test result: the
measured value displays and flashes.
Leak detection:
Memo function
Setting
Function activation.
Time delay period
Display time delay activation:
On : the measured leak value will flash during the time delay
period set.
Off : the measured leak value will flash until a new test will be
started.
•Quickaccesstosettingfromthecontrolpanel:
Leak detection:
Zero activation
function
This function allows to help the operator to identify very small He
signal variations in the surrounding background or to dilate small
measured signal fluctuations on the bargraph display, for example at
the pump slowing down.
Function triggering mode
Function triggering set point
If "auto, to set
Activated function if "automatic"
Manual activation of the Zero function: .
•ItisadvisedtousethisfunctionwhentheHebackgroundishigh
andstable.
•Thisfunctionallowstomeasurealeakow2rangesbelowthe
detectorbackgroundinprogress,assoonasthisoneisnomorein
roughing.
adixen Vacuum Products - ASM 310 Operating instructions