Start-up
Switching off
recommanded
The leak detector can be switched off at any time by depressing the
main power switch to 0 (OFF) but it is advised (especially when the
leak detector may not be used for an extended period of time) to
stop it following the procedure described below.
Place the blank-off flange or close the test chamber.
Start a test cycle.
Wait until the leak detector reaches the more sensitive test mode
Make sure that the inlet vent is OFF.
Stop the test cycle.
Stop the leak detector.
Wait before moving the detector so as to make sure of the
complete turbomolecular pump stop.
The leak detector is so under vacuum avoiding an important
degassing time in a future use.