4/5
C 402
Test Menu
GB 03341 - Edition 03 - March 2010
This function allows an automatic air inlet at the end of a test in
hard vacuum test.
This function allows to put back to atmospheric pressure the
detector inlet and so the part or volume connected.
This function is secure: a confirmation request displays each time
that the operator requests an air inlet.
Air inlet
Inlet vent is automatically done when the operator press
START
STAND-BY
to stop
the test if "Automatic" is selected.
If "Operator" is selected, it is necessary to press the dedicated
function key to put back the inlet detector to atmosphere.
Intet vent manual activation:
From thumb
(if parameted) or From standard window
Setting
If "automatic", set it
Optional automatic closing function
Time between the valve opening and its automatic
closing after.
Activated function if "Automatic".
Time between the test stop and the
valve opening.
Optional automatic
closing activation.
Inlet Vent
adixen Vacuum Products - ASM 310 Operating instructions