2/4
C 500
Factory configuration of the leak detector
parameters
GB 03291 - Edition 02 - January 09
Parameters
Configuration
Factory
Customer
modification
Method
Vacuum
Mode
High
sensitivity
Hard vacuum correction
off
Hard vacuum correction value
1
.00E+00
Sniffing correction
off
Sniffing correction value
1.0E+00
Cycle end
Operator
Roughing time activated
yes
Roughing time
00:
1
0
Measure time
00:1
0
Inlet vent
Operator
Delay
0 s
Time activation
off
Time
10 s
Tracer gas
He
Location
Internal
Calibration valve
Closed
Internal calibrated leak value
See the
calibration
certificate of
the internal
calibrated
leak delivered
with the
detector
Calibration year
Unit
Loss per year (%/yr)
T° Coeff. (%/°C) :
Ref. T°
Primary pump maintenance counter
initial value
10000
Secondary pump maintenance counter
initial value
15000
Cycles maintenance counter initial
value
5.00E-05
C 402
C 403
C 404
adixen Vacuum Products - ASM 310 Operating instructions