EasyManua.ls Logo

Agilent Technologies 1260 - Optimization Overview

Agilent Technologies 1260
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity DAD and MWD User Manual 117
How to optimize the Detector
6
Optimization Overview
Optimization Overview
Tab le 27 Optimization Overview
Parameter Impact
1 Selection of flow cell peak resolution versus sensitivity
Choose flow cell according to used column, see Figure 50 on page 118.
2 Connection of flow cell chromatographic resolution
For flow rates from 0.5 ml/min connect column using the zero-dead-volume
fittings of the detector.
For small column i.d. (e.g 1 mm) the inlet capillary of the micro flow cell can
be connected directly to the column.
3 Setting the peak width (response time) peak resolution versus sensitivity
versus disk space
Use peak width according Figure 50 on page 118 as starting point.
Set the peak-width close to the width of a narrow peak of interest in your
chromatogram.
4 Setting wavelength and bandwidth
Sample wavelength:
Never miss a peak by the use of a browser wavelength like 250 nm with
100 nm bandwidth.
Select specific wavelength with reduced bandwidth if you need
selectivity, e.g. 250,10 nm and 360,100 nm as reference wavelength.
Set the sample wavelength to a peak or valley in the spectrum to get
best linearity for high concentrations.
sensitivity versus selectivity
sensitivity versus linearity
Reference wavelength:
Select the reference wavelength with broad bandwidth (30...100 nm)
wavelength range where your analytes have little or no absorbance (e.g.
sample at 254 nm, reference at 320 nm).
baseline drift due to RI effects.

Table of Contents

Other manuals for Agilent Technologies 1260

Related product manuals