Error Information
8
Detector Error Messages
UV Ignition Failed
The UV lamp failed to ignite.
The processor monitors the UV lamp current during the ignition cycle. If the
lamp current does not rise above the lower limit within 2 5 seconds, the
error message is generated.
Probable cause Suggested actions
1 Lamp too hot. Hot gas discharge lamps may
not ignite as easily as cold lamps.
Switch off the lamp and allow it to cool down for
at least 15 minutes.
2 Lamp disconnected.
Ensure the lamp is connected.
3 Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
4 Defective detector main board.
Please contact your Agilent service
representative.
5 Defective power supply.
Please contact your Agilent service
representative.