Agilent 1260 Infinity DAD and MWD User Manual 53
Installing the Module
3
Flow Connections to the Detector
Flow Connections to the Detector
Parts required # Description
G1315-68755 Accessory kit
Hardware required Other modules
Preparations Detector is installed in the LC system.
WARNING
Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
➔ When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor and follow good
laboratory practice.
➔ The amount of substances should be reduced to the minimal volume required for
the analysis.
➔ Do not operate the instrument in an explosive atmosphere.
NOTE
The flow cell is shipped with a filling of isopropanol (also recommended when the
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
NOTE
The detector should be operated with the front cover in place to protect the flow cell area
against strong drafts from the outside and to cover the deuterium lamp.
Some types of the Agilent deuterium lamps show a light ring during operation. This is not
harmful, refer to “UV-Radiation” on page 272.