EasyManuals Logo
Home>Agilent Technologies>Security Sensors>1260

Agilent Technologies 1260 User Manual

Agilent Technologies 1260
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #195 background imageLoading...
Page #195 background image
Agilent 1260 Infinity DAD and MWD User Manual 195
Maintenance
10
Overview of Maintenance
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector that
can be carried out without opening the main cover.
Tab le 34 Overview of Maintenance
Procedure Typical Frequency Notes
Cleaning of module If required.
Deuterium lamp or
tungsten lamp
exchange
If noise and/or drift exceeds your application limits or
lamp does not ignite.
An intensity test should be performed
after replacement.
Flow cell exchange If application requires a different flow cell type. A holmium or wavelength calibration
test should be performed after
replacement.
Flow cell parts
Cleaning or exchange
If leaking or if intensity drops due to contaminated flow
cell windows.
A pressure tightness test should be
done after repair.
Holmium oxide filter
Cleaning or exchange
If contaminated. A holmium or wavelength calibration
test should be performed after
replacement.
Leak sensor drying If leak has occurred. Check for leaks.
Leak handling System
replacement
If broken or corroded. Check for leaks.

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals