38 Agilent 1260 Infinity DAD and MWD User Manual
2 Site Requirements and Specifications
Performance Specifications
Performance Specifications
Specifications
Tab le 13 Performance Specifications G1315C/D and G1365C/D
Type Specification Comments
Detection type 1024-element photodiode array
Light source Deuterium and tungsten lamps The UV-lamp is equipped with RFID tag that
holds lamp typical information.
Data rate up to 80 Hz (G1315C/G1365C)
up to 20 Hz (G1315D/G1365D)
Wavelength range 190 – 950 nm
Short term noise
(ASTM) Single and
Multi-Wavelength
< ± 0.7·10
-5
AU at 254 and 750 nm
see "Specification Conditions" below
Drift
< 0.9·10
-3
AU/h at 254 nm
see "Specification Conditions" below
Linear absorbance
range
> 2 AU (5 %) at 265 nm see "Specification Conditions" below
Wavelength accuracy ± 1 nm Self-calibration with deuterium lines, verification
with holmium oxide filter
Wavelength bunching 1 – 400 nm Programmable in steps of 1 nm
Slit width 1, 2, 4 , 8, 16 nm Programmable slit
Diode width < 1 nm