76 Agilent 1260 Infinity DAD and MWD User Manual
4 LAN Configuration
Storing the settings permanently with Bootp
Storing the settings permanently with Bootp
If you want to change parameters of the module using the Bootp follow the
instructions below.
1 Turn off the module.
2 Change the module's settings of the Configuration Switch to “Bootp &
Store” mode, see Table 16 on page 61.
3 Start the Agilent Bootp Service and open its window.
4 If required, modify the parameters for the module according to your needs
using the existing configuration.
5 Press OK to exit the Bootp Manager.
6 Now turn on the module and view the Bootp Server window. After some
time the Agilent Bootp Service will display the request from the LAN
interface. The parameters are now stored permanently in the non-volatile
memory of the module.
7 Close the Agilent Bootp Service and turn off the module.
8 Change the settings of the module’s Configuration Switch to “Using Stored”
mode, see Table 16 on page 61.
9 Power cycle the module. The module can be accessed now via LAN without
the Agilent Bootp Service.

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals