UV-Radiation
UV-Radiation
Emissions of ultraviolet radiation (200-315 nm) from this product is limited
such that radiant exposure incident upon the unprotected skin or eye of
operator or service personnel is limited to the following TLVs (Threshold Limit
Values) according to the American Conference of Governmental Industrial
Hygienists:
Typically the radiation values are much smaller than these limits:
NOTE
This information is only valid for UV-lamps without cover (e.g. 2140-0590 and 2140-0813).
Tab le 38 UV-Radiation Limits
Exposure/day Effective Irradiance
8hours 0.1µW/cm2
10 minutes 5.0 µW/cm2
Tab le 39 UV-Radiation Typical Values
Position Effective Irradiance
Lamp installed, 50 cm distance Average 0.016 µW/cm2
Lamp installed, 50 cm distance Maximum 0.14 µW/cm2