Agilent 1260 Infinity DAD and MWD User Manual 209
Maintenance
10
Replacing Capillaries on a Standard Flow Cell
Replacing Capillaries on a Standard Flow Cell
When If the capillary is blocked
Tools required Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
Parts required Description
For parts see “Standard Flow Cell” on page 234.
Preparations Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see “Exchanging a Flow Cell” on page 200.
NOTE
All descriptions in this procedure are based on the default orientation of the cell (as it is
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
NOTE
The fittings at the flow cell body are special types for low dead volumes and not compatible
with other fittings.
When retightening the fittings, make sure that they are carefully tightened (handtight plus
1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals